Description | Glass filter equivalent to NIST SRM 2065 |
---|---|
Primary Usage | Validation of the wavelength scale over wide range |
Useable Range | 2335nm to 1945nm |
Physical Configuration | Glass filter with aluminum housing |
NIST Traceability | Each set is supplied with a Certificate of Analysis and NIST Traceability |
ISO Accredited | Accredited by UKAS ISO 17025, ISO Guide34 and ISO 9001 |
This exceptionally wide-range filter contains a mixture of rare earth oxides and provides 20 peaks for wavelength calibration between 335 nm in the UV and 1945 nm in the NIR. It is equivalent to NIST SRM 2065.
Nominal peak wavelength values (in nm) are: 335, 345, 361, 375, 386, 402, 418, 485, 538, 583, 642, 748, 804,976, 1076, 1151, 1222, 1366, 1469, 1945
Catalog Number | Description |
---|---|
RM-2065 | Starna wide range wavelength reference |