Description | All of the references in one set for qualification of UV/VIS spectrophotometer for USP |
---|---|
Primary Usage | Validate the performance of UV spectrophotometers for USP qualification |
Useable Range | See lising of references in each set below |
Physical Configuration | Far UV quartz cells that have been permanently sealed by heat fusion and glass filters in aluminum holders |
NIST Traceability | Each set is supplied with a Certificate of Analysis and NIST Traceability |
ISO Accredited | Accredited by UKAS ISO 17025, ISO Guide34 and ISO 9001 |
The United States Pharmacopeia Chapter <857> (May 2015) gives new recommendations for the qualification of ultraviolet and visible spectrophotometers, including tests for:
Absorbance accuracy and linearity
Wavelength accuracy
Stray light (new test procedure)
Resolution (new qualification requirement)
The new procedures became mandatory from 1st May 2016. These Starna Qualification Sets contain a number of Certified Reference Materials required to perform these tests.
This set contains Starna Certified Reference Materials to qualify a UV spectrophotometer to USP Chapter <857>
Potassium Dichromate 60 mg/l & blank cell | Photometric Accuracy, 235 - 350 nm |
Holmium Oxide Solution cell | Wavelength Accuracy, 240 - 640 nm |
Potassium Chloride 10 & 5 mm cells | Stray Light, 190 - 205 nm |
Sodium Iodide 10 & 5 mm cells | Stray Light, 210 - 259 nm |
Acetone 10 & 5 mm cells | Stray Light, 250 - 320 nm |
Sodium Nitrite 10 & 5 mm cells | Stray Light, 300 - 385 nm |
Toluene in Hexane & blank cell | Resolution, 267 - 269 nm |
This set is the same as RM-USP, but contains two additional potassium dichromate references for checking photometric linearity.
Potassium Dichromate 60 mg/l & blank cell | Photometric Accuracy, 235 - 350 nm |
Potassium Dichromate 20 mg/l | Photometric Accuracy, 235 - 350 nm |
Potassium Dichromate 100 mg/l | Photometric Accuracy, 235 - 350 nm |
Holmium Oxide Solution cell | Wavelength Accuracy, 240 - 640 nm |
Potassium Chloride 10 & 5 mm cells | Stray Light, 190 - 205 nm |
Sodium Iodide 10 & 5 mm cells | Stray Light, 210 - 259 nm |
Acetone 10 & 5 mm cells | Stray Light, 250 - 320 nm |
Sodium Nitrite 10 & 5 mm cells | Stray Light, 300 - 385 nm |
Toluene in Hexane & blank cell | Resolution, 267 - 269 nm |
For method validation, USP <857> recommends that absorbance linearity is checked with five or more analyte concentrations. The following Potassium Dichromate sets allow instrument linearity to be checked over the relevant absorbance range.
Photometric Linearity Set, 0.095A -1.5A | Part Number RM-0204060810 |
Potassium Dichromate 20, 40, 60, 80, 100 mg/l & blank cell | |
Photometric Linearity Set, 1.5A- 3.3A | Part Number RM-121416182224 |
Potassium Dichromate 120, 140, 160, 180, 220 & 240 mg/l & blank cell | |
Photometric Linearity Set, 0.19A - 3.3A | Part Number RM-040812162024 |
Potassium Dichromate 40, 80, 120, 160, 200 & 240 mg/l & blank cell |
These references are recommended in USP Chapter <857> for the qualification of spectrophotometers in the Visible region:
Absorbance Accuracy, 440-635 nm | Neutral Density Filters (1.00, 0.50 & 0.25 A) |
Wavelength Accuracy, 240 - 640 nm | Holmium Glass Filter |
These SRMs are specifically cited in USP <857> as evidence of NIST traceability.
SRM 930e Set | Part Number: RM-1N2N3N |
Absorbance Accuracy, 440-635 nm Neutral Density Filters: 10, 20 & 30 %T | |
SRM 1930 Set | RM-N1N35N |
Absorbance Accuracy, 440-635 nm Neutral Density Filters: 1, 3 & 50 %T | |
SRM 2930 Set | RM-D1D39N |
Absorbance Accuracy, 440-635 nm Neutral Density Filters 0.1, 0.3 & 92 %T | |
Combined SRM Set | RM-SRM9ND |
Absorbance Accuracy, 440-635 nm All 9 Neutral Density Filters listed above |